东芝 spellman电源维修CT42 CT70北京
SPELLMAN高压电源广泛应用于东芝 XVISION/EX、AUKLET系列、ASTEION系列、以及多排系列。故障代码及诊断方法大致相同,部分故障代码排查参考如下
Error contents (note 1) | Cause of problem | Object part | ||
24kW | 48kW/60kW | |||
E13 | Arc- | Cathode side Arc of the X-ray tube or the high-voltage section | TUBE, INV/HV | TUBE, INV/HVC |
E25 | Arc+ | Anode side Arc of the X-ray tube or the high-voltage section | TUBE, INV/HVA | |
E12 | Over kV- | Cathode side overvoltage. Abnormal main circuit control. | INV/HV, SCB | INV/HVC, SCB |
E27 | Over kV+ | Anode side overvoltage. Abnormal main circuit control. | INV/HVA, SCB | |
E62 | Over mA- | Cathode side overcurrent of the X-ray tube or the high-voltage section. | TUBE, INV/HV, SCB | TUBE, INV/HVC,SCB |
E63 | Over mA+ | Anode side overcurrent of the X-ray tube or the high-voltage section. | TUBE, INV/HVA,SCB | |
E66 | MAC error | Filament current abnormal. Trouble of the filament heatig circuit. | INV/HV, SCB | INV/HVC, SCB |
E0 | Regulation error (HV abnormal) | There is a difference in binding output and real output | INV/HV, AC/C | INV/HVC, INV/HVA, AC/C |
spellman电源维修 高压电源维修CT70PN50X4218北京
医疗设备高压电源维修高压发生器维修北京天浦正达高压电源专业维修
Spellman高压电源XRM/PCM50/70维修射频电源北京
Spellman高压电源XRM50P50X3768维修射频电源北京
spellman高压电源维修高压发生器控制单元维修北京
医疗设备高压电源维修高压发生器维修北京天浦正达高压电源专业维修